3CALCulate Subsystem
94 U2000 Series Programming Guide
CALCulate[1]:LIMit:FCOunt?
This query returns the total number of limit failures for the current
window/measurement.
If the appropriate STATe commands are set to ON, each time a measurement is
initiated on the current window/measurement and the result is outside the
limits, the counter is incremented by one.
If the measured value is equal to a limit, this is a limit pass.
The counter is reset to zero by any of the following commands:
• *RST
• CALCulate[1]:LIMit:CLEar:IMMediate
• CALCulate[1]:LIMit:CLEar:AUTO ON
When CALCulate[1]:LIMit:CLEar:AUTO is set to ON, the counter is set to
zero each time a measurement is:
•measured using MEASure?
•read using READ?
• initiated using:
• INITiate[:IMMediate] or,
• INITiate:CONTinuous ON
When CALCulate[1]:LIMit:CLEar:AUTO is set to ONCE, the counter is set to
zero the first time a measurement is:
•measured using MEASure?
•read using READ?
• initiated using:
• INITiate[:IMMediate] or,
• INITiate:CONTinuous ON
The maximum number of errors is 2
16
–1. If more than 2
16
–1 errors are
detected the counter returns to zero.

Table of Contents

Other manuals for Agilent Technologies U2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U2000 Series and is the answer not in the manual?

Agilent Technologies U2000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelU2000 Series
CategoryAccessories
LanguageEnglish

Related product manuals