7SENSe Subsystem
198 U2000 Series Programming Guide
SENSe[1]:POWer:AC:RANGe <numeric_value>
Its purpose is to select one of two power ranges.
If 0 is selected, the power sensor’s lower range is selected
If 1 is selected, the power sensor’s upper range is selected
Setting a range with this command automatically switches
SENSe[1]:POWer:AC:RANGe:AUTO to OFF.
Syntax
Example
Reset Condition
On reset, the upper range is selected.
Query
SENSe[1]:POWer:AC:RANGe?
The query enters a 1 or 0 into the output buffer indicating the status of the
power sensor’s range.
POW:AC:RANG 0 This command sets the power sensor to
it’s lower range.
SENS
1 :AC
POW
:
:RANG
Space
?
numeric_value