EasyManuals Logo

Agilent Technologies U2000 Series Guide

Agilent Technologies U2000 Series
335 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #290 background imageLoading...
Page #290 background image
11 TRIGger Subsystem
270 U2000 Series Programming Guide
INITiate[1]:CONTinuous:SEQuence[1] <boolean>
This command sets the power sensor for either a single trigger cycle or
continuous trigger cycles. A trigger cycle means that the power sensor exits
the wait for trigger state and starts a measurement.
If INITiate:CONTinuous:SEQuence[1] <boolean> is set to:
• OFF, the trigger system remains in the idle state until it is set to ON, or
INITiate:IMMediate is received. Once this trigger cycle is complete the
trigger system returns to the idle state.
• ON, the trigger system is initiated and exits the idle state. On completion of
each trigger cycle, the trigger system immediately commences another
trigger cycle without entering the idle state.
Syntax
Example
NOTE
This command performs the same functions as INITiate[1]:CONTinuous
<boolean>.
INIT:CONT:SEQ1 ON This command places the sensor in a wait
for trigger state.
INIT
:CONT
0|OFF
1|ON
Space
?
:SEQ
1

Table of Contents

Other manuals for Agilent Technologies U2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U2000 Series and is the answer not in the manual?

Agilent Technologies U2000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelU2000 Series
CategoryAccessories
LanguageEnglish

Related product manuals