TRIGger Subsystem 11
U2000 Series Programming Guide 281
TRIGger[:SEQuence]:SLOPe <character_data>
This command specifies whether a trigger event is recognized on the rising or
falling edge of a signal.
Syntax
Parameters
Reset Condition
On reset the value is set to POSitive.
Item Description/Default Range of Values
character_data How a trigger event is recognized:
• POSitive: a trigger event is recognized on
the rising edge of a signal.
• NEGative: a trigger event is recognized on
the falling edge of a signal.
POSitive
NEGative