TRIGger Subsystem 11
U2000 Series Programming Guide 283
TRIGger[:SEQuence[1]]:COUNt <numeric_value>
This command controls the path of the trigger subsystem in the upward
traverse of the wait for trigger state. COUNt loops through the event
detection/measurement cycle are performed. That is, COUNt measurements
are performed in response to COUNt trigger events.
COUNt can be set to a value >1 only when:
• [SENSe[1]]:MRATe <character_data> is set to FAST
and
• TRIGger:SOURce set to IMMediate or HOLD.
Syntax
Parameters
Item Description/Default Range of Values
numeric_value The number of triggered events for the
measurement cycle.
• DEF: the default value is 1
1 to 50
DEF
:COUN Space
?
numeric_value
DEF
TRIG
:SEQ
1