1 Power Sensor Remote Operation
54 U2000 Series Programming Guide
Using “*” Commands
Commands starting with a “*” are called common commands. They are
required to perform the identical function for all instruments that are
compliant with the IEEE- 488.2 interface standard. The “*” commands are
used to control reset, self- test, and status operations in the power sensor.
Syntax Conventions
Throughout this guide, the following conventions are used for SCPI command
syntax.
Square brackets ([]) indicate optional keywords or parameters.
Braces ({}) enclose one or more parameters that may be included zero or
more times.
Triangle brackets (<>) indicate that you must substitute a value for the
enclosed parameter.
Bars (|) can be read as “or” and are used to separate alternative parameter
options.
Syntax Diagram Conventions
Solid lines represent the recommended path.
Ovals enclose command mnemonics. The command mnemonic must be
entered exactly as shown.
NOTE
If you send two query commands without reading the response from the first, then attempt
to read the second response, you may receive some data from the first response followed
by the complete second response. To avoid this, do not send a query command without
reading the response. When you cannot avoid this situation, send a device clear before
sending the second query command.

Table of Contents

Other manuals for Agilent Technologies U2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U2000 Series and is the answer not in the manual?

Agilent Technologies U2000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelU2000 Series
CategoryAccessories
LanguageEnglish

Related product manuals