11 TRIGger Subsystem
288 U2000 Series Programming Guide
TRIGger[:SEQuence[1]]:SOURce
BUS|EXTernal|HOLD|IMMediate
This command configures the trigger system to respond to the specified
source. This command only selects the trigger source. Use the INITiate
command to place the power sensor in the wait for trigger state.
Syntax
NOTE
This command has the same purpose as TRIGger[1]:SOURce
BUS|EXTernal|HOLD|IMMediate.
:SOUR
TRIG
:SEQ
1
EXT
HOLD
Space
IMM
BUS
?