TRIGger Subsystem 11
U2000 Series Programming Guide 265
INITiate[1]:CONTinuous <boolean>
This command sets the power sensor for either a single trigger cycle or
continuous trigger cycles. A trigger cycle means that the power sensor exits
the wait for trigger state and starts a measurement.
If INITiate:CONTinuous is set to:
• OFF, the trigger system remains in the idle state until it is set to ON, or
INITiate:IMMediate is received. Once this trigger cycle is complete the
trigger system returns to the idle state.
• ON, the trigger system is initiated and exits the idle state. On completion of
each trigger cycle, the trigger system immediately commences another
trigger cycle without entering the idle state.
Syntax
NOTE
This command performs the same function as INITiate:CONTinuous:SEQuence[1]
<boolean>.
INIT
1
:CONT
0|OFF
1|ON
Space
?