TRIGger Subsystem 11
U2000 Series Programming Guide 265
INITiate[1]:CONTinuous <boolean>
This command sets the power sensor for either a single trigger cycle or
continuous trigger cycles. A trigger cycle means that the power sensor exits
the wait for trigger state and starts a measurement.
If INITiate:CONTinuous is set to:
• OFF, the trigger system remains in the idle state until it is set to ON, or
INITiate:IMMediate is received. Once this trigger cycle is complete the
trigger system returns to the idle state.
• ON, the trigger system is initiated and exits the idle state. On completion of
each trigger cycle, the trigger system immediately commences another
trigger cycle without entering the idle state.
Syntax