6MEMory Subsystem
150 U2000 Series Programming Guide
Example
Query
MEMory:TABLe:FREQuency?
The query returns a list of frequency points for the table currently selected.
The frequencies are returned in Hz.
Query Example
Error Messages
If more than 80 frequencies are in the list, error –108, “Parameter not
allowed” occurs.
If the frequencies are not entered in ascending order, error –220,
“Parameter error;Frequency list must be in ascending order” occurs.
If a table has not been specified using the MEMory:TABLe:SELect
command, the data cannot be entered into the table and error –221,
“Settings conflict” occurs.
If a frequency is set which is outside of the allowed frequency range, error
–222, “Data out of range” occurs.
MEM:TABL:FREQ
200kHz,600kHz
This command enters frequencies of 200
kHz and 600 kHz into the currently
selected table.
MEM:TABL:FREQ? This command queries the frequency
points in the currently selected table.

Table of Contents

Other manuals for Agilent Technologies U2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U2000 Series and is the answer not in the manual?

Agilent Technologies U2000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelU2000 Series
CategoryAccessories
LanguageEnglish

Related product manuals