EasyManuals Logo

Agilent Technologies U2000 Series Guide

Agilent Technologies U2000 Series
335 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
1 Power Sensor Remote Operation
6 U2000 Series Programming Guide
command CALibration[1]:AUTO ONCE shows the code compatibility with
CALibration[1]:ZERO:AUTO ONCE. Users are not required to modify their
source code which is used in other power meters or sensors as
CALibration[1]:AUTO ONCE supports backward compatibility.
Calibration Sequence
This feature allows you to perform a complete calibration sequence with a
single query. The query is:
CALibration[1][:ALL]?
The complete calibration sequence consists of:
Zeroing and calibrating the power sensor (CALibration[1]:ZERO:AUTO
ONCE or CALibration[1]:AUTO ONCE)
The query enters a number into the output buffer when the sequence is
complete. If the result is 0 the sequence was successful. If the result is 1 the
sequence failed. Refer to “CALibration[1][:ALL]?” on page 110 for further
information.
NOTE
The CALibration[1][:ALL] command is identical to the
CALibration[1][:ALL]? query except that no number is returned to indicate the
outcome of the sequence. You can examine the Questionable Status Register or the error
queue to discover if the sequence has passed or failed. Refer to “Status Reporting” on
page 36 for further information.

Table of Contents

Other manuals for Agilent Technologies U2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U2000 Series and is the answer not in the manual?

Agilent Technologies U2000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelU2000 Series
CategoryAccessories
LanguageEnglish

Related product manuals