EasyManua.ls Logo

Anritsu ML2437A - Page 128

Anritsu ML2437A
314 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Example 11: Frequency Measurements Using a Separate Source Measurement Examples
7-32 PN: 10585-00001 Rev. P ML2437A/38A OM/PM
Press [START] and using the keypad, input the frequency at which the sweep starts.
Press [Enter] to accept the value.
Press [STOP] and using the keypad, input the frequency at which the sweep stops.
Press [Enter] to accept the value.
The following screen is displayed:
Note that the frequency axis reflects the Start and Stop settings. Each time you
measure at different frequencies, it is necessary to re-input the [START] and [STOP]
values.
To achieve the maximum dynamic range, apply the highest levelled output power for
the source as possible i.e. +20 dBm.
3. Changing the power axis
To change the power axis, press [System]
> [Control] > [-more-] > [SCALE].
Press [TOP] and/or [BOTTOM] to set the dB values as required. Press [Enter] to accept
the values.
In Source Sweep mode, you are unable to measure levels of less than -50 dBm.
In Source Sweep mode, the calibration factor is handled in a different manner since it is
defined by the current frequency of the source. Press [Sensor]
> [Cal Factor].
Press [SOURCE] and toggle the setting by repeatedly pressing the soft key until the
V/GHz is displayed.
To display the Measurement screen, press any hard key.
Figure 7-48.
Figure 7-49.

Table of Contents

Related product manuals