Connecting the TDR9000
Physical Front Panel
4-38 72A-1898 Rev. A 11/01
Monday, November 26, 2001 1:09 pm
To set up a current shunt:
1. Connect the Instrument end of the Doble supplied Analog cable to a
pair of Analog channel banana jacks.
The Analog Channel inputs are polarity sensitive.
2. Make sure to attach the shield (green/yellow) banana plug to the
chassis ground (green/yellow) banana jack on the Event module.
Multiple shields can be stacked (see Figure 4.13 on page 4-30).
3. Taking appropriate safety precautions, use the attachment accessories
provided to attach the other end of the cable across the device to be
monitored, either directly across the device or at the terminal block.
For best results, Doble recommends that the shunt be installed on the
low or neutral side of the circuit through which the test current flows.
System Module
Trigger In/Out
Setup
The Trigger In and Trigger Out connections on the System module are
connected as follows:
1. Attach one end of the Doble supplied Auxiliary Contact cable to a
pair of Trigger In or Trigger Out channel banana jacks (K on
Figure 4.13 on page 4-30).
2. Taking appropriate safety precautions, use the attachment accessories
provided to attach the other end of the cable across the device to be
monitored, either directly across the device or at the terminal block.
N
OTE
Trigger Out performs no monitoring on the device described in step 2.
WARNING The System module Trigger In channel is the only TDR9000 channel
rated to 600 V peak. When connecting this channel use appropriately
rated test leads.