DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
4-18 Section 4-3 - General Procedure
4-3-10-1 Using the DVD-R (cont’d)
15.)It will do Verifying.
16.)It will eject the CD/DVD and Go to Utility and reinsert the CD/DVD.
17.)Go to Utility->Connectivity->Removable Media.
18.)Press Verify
19.)The properties will be like below.
a.) Capacity is the size of transferred data.
b.) Free space is 0.
c.) Formatted / Database Present / Finalized : Yes
(DICOMDIR Present : depends on setting)
Figure 4-12 Verifying
Figure 4-13 Removable Media
NOTICE
Avoid mechanical ejection whenever possible. Mechanical ejection leaves the actuator
unlocked. If forced to use this method, reboot the system, then insert and eject a known good
disk using one of the other methods.

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
Power Supply100-240VAC, 50/60Hz
Display15" LCD
Doppler ModesColor Doppler, Power Doppler
Imaging ModesB-Mode, M-Mode
ProbesConvex, Linear, Phased Array
ApplicationsAbdominal, Vascular, Cardiac, OB/GYN, Small Parts, Musculoskeletal
ConnectivityDICOM, USB, Ethernet
Probe Connectors3
PortabilityMobile
Operating Temperature10°C to 40°C
Relative Humidity30%

Related product manuals