GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 4-9 - Board Function Checks 4-51
4-9-9 BackPlane Assy Function Validation Procedure
1.) Turn on the system
2.) All of attached ASSY should be tested.
3.) Refer to the 4-9-4 "SYSCONML Assy Function Validation Procedure" on page 4-47.
4.) Refer to the 4-9-7 "SOM Function Validation Procedure" on page 4-50.
5.) Refer to the 4-9-1 "CL1TRX Assy / BL1TRX Assy Function Validation Procedure" on page 4-41.
6.) Refer to the 4-9-6 "P3RLY,P2RLY and P2RLY with Dummy Assy function check procedure" on page
4-49.
7.) Refer to 4-9-5 "APS/APS Pro Assy PCB Function Validation Procedure" on page 4-48.
8.) Refer to 4-9-2 "ACWD Assy Function Validation Procedure" on page 4-42.(Optional)