4-3-11 Backup and Restore Database, Preset Configurations and Images
NOTE: When upgrading from R1.0.x or R2.0.x to R3.0.x or R4.0.x ,preset cannot be transferred. Refer
to the table below.
4-3-11-1 Formatting Media
1.) To format the backup media, CD-R or DVD-R, select the UTILITY key on the Front Panel.
2.) Select CONNECTIVITY
, then RERNOVABLE MEDIA. Properly label and Insert the backup
media.
3.) Select the media type from the drop down menu.
4.) Enter the label for the media as shown in Figure 4-14. It is best to use all capital letters with no
Table 4-4 LP5 BT11(R4.x.x) compatibility
Preset
Compatibility
R1.x.x R2.x.x R3.x.x
Patient Archive
Yes Yes Yes
Report Archive
Yes Yes Yes
Imaging Presets
No No Yes
Connectivity Configuration
No No Yes
Measurement Configuration
No No Yes
Comment/Body Pattern Libraries
No No Yes
Report Templates (Same Software Version Only)
No No No
3D/4D
No No Yes
Key Macro
No No Yes
All Others
No No Yes