6-4 Section 6-3 - Monitor Adjustments
6-3-3 LCD Plastic filter
6-3-3-1 Tools
• Common Phillips screwdrivers
• Allen/Unbraco wrench
• Stubby screwdriver (Flat tip and Cross tip)
6-3-3-2 Removal procedure
1.) Remove the Monitor cover set. Refer to the 8-2-2 "Monitor cover set" on page 8-5 and the figure
below.
2.) Remove the LCD Plastic filter.
3.) Put the LCD plastic filter on the LCD monitor after detaching the tape of the LCD plastic filter.
Figure 6-3 Removing the LCD plastic filter
Figure 6-4 Putting the LCD plastic filter on the LCD monitor