DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 4-7 - Keyboard Function Checks 4-33
4-7-6 TGC key assay / TGC Knob Set validation
1.) Follow up the procedure 4-7-1 "Main Keyboard Assy Function Validation Procedure" on page 4-29
to run the scfpapitest program.
2.) Slide TGC pot and check up the value.
4-7-7 Sub Keyboard Encoder Knob Set validation
Follow up 4-7-2 "Sub Keyboard Assy Function Validation Procedure" on page 4-32.
NOTICE
While sliding TGC pot, TGC message should be displayed.