GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
4-30 Section 4-7 - Keyboard Function Checks
4-7-1 Main Keyboard Assy Function Validation Procedure (cont’d)
6.) Write the 'scfpapitest.exe' on the command prompt and press 'Enter' button to pop up the
scfpapitest program.
7.) Press the keyboard button and rotate rotary button. When you press or rotary keys, Log dialog box
displays the key log. Button name and displaying log name is like the table.
Figure 4-29 scfpapitest program
Table 4-8 Button and Rotaty Table
Button or Rotary Name The Name displayed in Log Button or Rotary Name The Name displayed in Log
Patient
New Patient Button
Active Mode Active Mode Button
Probe/Preset
Probe Button
CF CFM PushEncoder Button
Report
Report Button
PDI PDI Button
End Exam
End Exam Button
Dept (Button) Depth PushEncoder Button
Record
Record button
Dept (Rotary) Depth PushEncoder Rotary
P1
Print or record button1
PW PWD PushEncoder Button
P2
Print or record button2
CW CW Button
Reverse
Reverse Button
M M Mode PushEncoder
Userdefine 1
Userdefine 1 Button
Scan Area ScanArea Button
Userdefine 2
Userdefine 2 Button
M/D Cursor MD Cursor Button
Userdefine 3
Userdefine 3 Button
Zoom (Button) Zoom/Size PushEncoder Button
Userdefine 4
Userdefine 4 Button
Zoom (Rotary) Zoom/Size PushEncoder Rotary
Userdefine 5
Userdefine 5 Button
Measure Measure Button
L (Left)
Left Button
BodyPattern/Ellipse (Button)
BodyPattern/Ellipse PushEncoder
Button

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
Power Supply100-240VAC, 50/60Hz
Display15" LCD
Doppler ModesColor Doppler, Power Doppler
Imaging ModesB-Mode, M-Mode
ProbesConvex, Linear, Phased Array
ApplicationsAbdominal, Vascular, Cardiac, OB/GYN, Small Parts, Musculoskeletal
ConnectivityDICOM, USB, Ethernet
Probe Connectors3
PortabilityMobile
Operating Temperature10°C to 40°C
Relative Humidity30%