GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
8-166 Section 8-5 - Full Backup/Restore Procedure
8-5-4 Restore procedure
1.) From Figure 8-225, press “Restore” button.
2.) Rest of the procedure is the same with “Backup” procedure.
3.) After reboot,
Press “Patient”. Verify that Patient data are restored
Press “Utility”. Verify that Userdefined data are restored.
NOTICE
The system should be rebooted after finishing of Restore procedures.