GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
8-144 Section 8-4 - Software Installation
8-4-3 Installing Application Software
NOTE: First time SW installation with peripherals on can cause the BlueScreen.
NOTE: When connecting peripherals to the system, make sure to turn off the system first in order to avoid the
BlueScreen.
1.) Place the “Application Software” CD into the DVD drive.
2.) Press the Start button on the tool bar at the bottom of the LCD screen and select
Run”.
Enter “
G:\LoadSoftware” into the dialog box as shown below.
3.) You will see a Command (CMD) window open as shown in the Figure 8-198.
4.) Press the “Y” key twice and copying backup files starts. This takes a few minutes.
WARNINGWARNING
Base Image software should also be installed prior to the application.
Figure 8-197 Run Load Software
Figure 8-198 Command Window

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
Power Supply100-240VAC, 50/60Hz
Display15" LCD
Doppler ModesColor Doppler, Power Doppler
Imaging ModesB-Mode, M-Mode
ProbesConvex, Linear, Phased Array
ApplicationsAbdominal, Vascular, Cardiac, OB/GYN, Small Parts, Musculoskeletal
ConnectivityDICOM, USB, Ethernet
Probe Connectors3
PortabilityMobile
Operating Temperature10°C to 40°C
Relative Humidity30%

Related product manuals