EasyManua.ls Logo

GE LOGIQ P5

GE LOGIQ P5
532 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
4-52 Section 4-9 - Board Function Checks
4-9-10 Nest fan Function Validation Procedure
1.) After replace Nest fan assy, turn on the system. Refer to 3-5-1 "Power On/Boot Up" on page 3-11.
2.) Confirm the fan rotating with naked eye and check the fan operating by rotating sound. No strange
sound should be heard.
NOTE: Do not close the right EMI bracket and right plastic cover before check the fan operation.
4-9-11 Transbox Function Validation Procedure
1.) Turn on the system. Refer to 3-5-1 "Power On/Boot Up" on page 3-11 for power up sequence to
check Transbox Assy function. The system must show correct operation as described on sequence.
2.) Check the peripheral operation. Refer to Section 4-5 "Peripheral Checks" on page 4-26 to confirm
the peripheral operation. All peripheral must work correctly as described Section 4-5.
4-9-12 ACPC Assy set Function Validation Procedure
1.) Turn on the system. Refer to 3-5-1 "Power On/Boot Up" on page 3-11 for power up sequence to
check Transbox Assy function. The system must show correct operation as described on sequence.
2.) Check the peripheral operation. Refer to Section 4-5 "Peripheral Checks" on page 4-26 to confirm
the peripheral operation. All peripheral must work correctly as described Section 4-5.
Figure 4-50 Checking Nest fan rotating

Table of Contents

Other manuals for GE LOGIQ P5

Related product manuals