DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 4-8 - Mechanical Function Checks 4-35
4-8-2 LCD plastic filter Function Validation Procedure
Check if LCD plastic filter is fastened so that it won't move.
4-8-3 Air filter set Function Validation Procedure
Check if the air filter is placed deep inside.
Figure 4-31 Check the LCD Plastic Filter
Figure 4-32 Check the air filter