DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
4-20 Section 4-3 - General Procedure
spaces or punctuation marks.
5.) The Ultrasound system displays a pop-up menu, as shown in Figure 4-15. When the formatting
has been completed, press OK
to continue.
6.) If desired, verify that the format was successful by returning to
Utility > Connectivity >
Removable Media
and selecting VERIFY as shown in Figure 4-12.
Figure 4-14 Format and Verify Media
Figure 4-15 Format Successful Pop-up Menu
Verify Format
Format
Media
Selection
Media Label