EasyManua.ls Logo

GE LOGIQ P5

GE LOGIQ P5
532 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 4-3 - General Procedure 4-21
4-3-11-2 Backup System Presets and Configurations
NOTE: Always backup any preset configurations before a software reload. This ensures that if the presets need
to be reloaded, after the software update, they will be the same ones the customer was using prior to
service.
1.) Insert a formatted CD-R or DVD-R into the drive.
2.) Press UTILITY
key
3.) Select SYSTEM
. on the monitor screen
4.) On the monitor display, select BACKUP/RESTORE
.
5.) In the Backup list, select Patient Archive, Report Archive, User Defined Configuration.
6.) In the Media field, select DVD-R.
7.) Select BACKUP
.
The system performs the backup. As it proceeds, status information is displayed on the Backup/Restore
screen.
Figure 4-16 Backup/Restore Menu
Check here to
backup presets
and configurations

Table of Contents

Other manuals for GE LOGIQ P5

Related product manuals