Section 4-3 - General Procedure 4-21
4-3-11-2 Backup System Presets and Configurations
NOTE: Always backup any preset configurations before a software reload. This ensures that if the presets need
to be reloaded, after the software update, they will be the same ones the customer was using prior to
service.
1.) Insert a formatted CD-R or DVD-R into the drive.
2.) Press UTILITY
key
3.) Select SYSTEM
. on the monitor screen
4.) On the monitor display, select BACKUP/RESTORE
.
5.) In the Backup list, select Patient Archive, Report Archive, User Defined Configuration.
6.) In the Media field, select DVD-R.
7.) Select BACKUP
.
The system performs the backup. As it proceeds, status information is displayed on the Backup/Restore
screen.
Figure 4-16 Backup/Restore Menu
Check here to
backup presets
and configurations

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
Power Supply100-240VAC, 50/60Hz
Display15" LCD
Doppler ModesColor Doppler, Power Doppler
Imaging ModesB-Mode, M-Mode
ProbesConvex, Linear, Phased Array
ApplicationsAbdominal, Vascular, Cardiac, OB/GYN, Small Parts, Musculoskeletal
ConnectivityDICOM, USB, Ethernet
Probe Connectors3
PortabilityMobile
Operating Temperature10°C to 40°C
Relative Humidity30%

Related product manuals