GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-107
8-2-47-3 Removal procedure (cont’d)
4.) Unplug the SATA cable from the backplane on the bottom of backplane.
5.) Disconnect the SATA Power Cable and SATA Cable from HDD Assy after removing 2ea cable ties.
And unscrew 4screws to remove the HDD Fixture bracket.
Figure 8-144 Unplug the SATA Cable
Figure 8-145 Separate HDD and HDD Fixture Bracket
Cable Tie