DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
8-122 Section 8-3 - Software Loading Procedure
8-3-5-2 Image Management Guide
NOTE: Before upgrade to R4.X.X software, perform recommended image backup procedure.
Export/Import
Data/Images Between Systems - Use this to copy both patient data and images for
specified patient(s) from one system to another.
To move exams from one Ultrasound system to another, you need to export/import exam information.
NOTE: Both database information and images are exported. No data is deleted from the local archive
when exporting data.
Exporting Data
To export an exam(s) to a compatible Ultrasound system:
1.) Format the removable media (DVD or CD-ROM). Label the removable media. Answer Yes/OK
to the messages.Press Patient. Deselect any selected patient(s) in the search portion of the
Patient screen. Press Data Transfer (located at the upper, right-hand corner of the Patient
menu).
2.) Select Export task Button.
3.) The [From] combo box is not active. It displays Local Archive. The [To] combo box is active.
Select the type of removable media CD/DVD. Then please wait until the patient list is visible.
Figure 8-160 Exam Data Transfer
Figure 8-161 Task button
Figure 8-162 To Combo Box

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
Power Supply100-240VAC, 50/60Hz
Display15" LCD
Doppler ModesColor Doppler, Power Doppler
Imaging ModesB-Mode, M-Mode
ProbesConvex, Linear, Phased Array
ApplicationsAbdominal, Vascular, Cardiac, OB/GYN, Small Parts, Musculoskeletal
ConnectivityDICOM, USB, Ethernet
Probe Connectors3
PortabilityMobile
Operating Temperature10°C to 40°C
Relative Humidity30%

Related product manuals