DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 10-3 - Maintenance Task Schedule 10-3
Table 10-2 Customer Care Schedule
Service at Indicated Time Daily Monthly Quarterly
Per Facilities
QA Program
Notes
Clean Probes •* * or before each use
Clean Probe Holders
Clean Air Filter
more frequently depending on
your environment
Inspect AC Mains Cable
Mobile Unit Check Weekly
Inspect Cables and Connectors
Clean Console
Clean Monitor and Touch Panel
Inspect Wheels, Casters, brakes and
Swivel Locks
Mobile Unit Check Daily
Check Control Panel Movement
Mobile Unit Check Daily
Console Leakage Current Checks
also after corrective
maintenance
Peripheral Leakage Current Checks
also after corrective
maintenance
Surface Probe Leakage Current Checks
also after corrective
maintenance
Endocavity Probe Leakage
Current Checks
also after corrective
maintenance
Transesphongeal Probe Leakage
Current Checks
also after corrective
maintenance
Surgical Probe Leakage
Current Checks
also after corrective
maintenance
Measurement Accuracy Checks
also after corrective
maintenance
Probe/Phantom Checks
also after corrective
maintenance
Functional Checks
also after corrective
maintenance