8-6-2-1 Perform Software Reloading in the Service Platform (cont’d)
17.)If “Disruptive Mode Status” is disabled, Press “Yes” to enable disruptive mode.
For the remote reloading, customer has to agree to the diruptive mode. The following window
(Figure 8-221) pops up in the system screen and customer has to press “Yes” to enable Disruptive
mode.
18.)Press “System Shutdown”
NOTE: If the service dongle is inserted or the system is logged on to the maintenance mode, “System
shutdown” will not work. Please close programs and reboot the system manually.
Figure 8-221 Insite Notification
Figure 8-222 System Shutdown