Configure Settings
2. Enable the Critical Value Notifications setting to configure notifications to
display when configured critical value thresholds are met or prescribed critical
conditions are detected.
3. Enable the Use Custom Notification Text setting to configure custom
notification text in the text field.
4. Edit the default phrase Critical Test Result in the text field with a customized
phrase. The phrase displays on the screen during preview or review of acquired
ECG patient reports, when a critical value or condition is detected.
5. Select the Threshold Configurations tab to display the threshold configurations
for critical values.
6. Select the default critical value to change the threshold value.
The selected value is now editable.
7. Enter the threshold for the selected critical value as per the information in the
table.
Table 42: Threshold Critical Values
Critical Value Allowed Threshold Range Default Threshold Value
Adult High HR 101 to 300 140
Adult Low HR 0 to 100 40
Pediatric High HR 101 to 300 180
Pediatric Low HR 0 to 100 50
High QTc 441 to 1000 550