Cleaning and Disinfection
• Sodium Hypochlorite (NaOCl) 5% solution)
• Isopropyl alcohol 70% (m/m)
Additional Information
• GE Healthcare used Super Sani-Cloth
®
wipes during cleaning and disinfection
validation. This cleaning agent is not listed in preference to other available
cleaning agents which may perform satisfactorily.
• The cleaning and disinfection information is provided in accordance with ANSI/
AAMIST81, ISO 17664. The recommendations provided above have been validated
as capable of preparing non-sterile GE Healthcare MAC
â„¢
5 medical devices.
It remains the responsibility of the user to make sure that the cleaning and
disinfection are performed using appropriate equipment, materials, and personnel
and achieves the desired result. This requires validation and routine monitoring
of the process. Any deviation from the provided recommendations should be
properly evaluated for effectiveness and potential adverse consequences.
• All users should be qualified personnel with documented expertise, competency,
and training. Users should be trained on hospital policies and procedures along
with current applicable guidelines and standards.
• Users should utilize appropriate Personal Protective Equipment (PPE) when
cleaning and disinfecting devices in accordance with the Department of
Environmental and Occupational Health and Safety’s (OSHA) blood-borne
pathogen guidelines or equivalent.
274  MAC
â„¢
5 A4/MAC
â„¢
5 A5/MAC
â„¢
5 Lite 5864335-001-1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 5 A4 and is the answer not in the manual?

GE MAC 5 A4 Specifications

General IconGeneral
BrandGE
ModelMAC 5 A4
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals