Configure Settings
Field Action Description
Suppress
Abnormal
Statements
Enable or disable
this setting.
If this setting is enabled, no abnormal or borderline
interpretative statements generate or display on the
report when you do as follows:
• View a report in the preview and review screens
• View a stored report in the review screen
• Print a report
• Send a report to a configured destination.
If this setting is disabled, abnormal and borderline
interpretative statements display on the report.
Default value: Disabled
Screening
Mode
Enable or disable
this setting.
If this setting is enabled, the device runs the 12SL
algorithm in high-specificity mode, where you will
not see certain lower-acuity statements in the
interpretation.
If this setting is disabled, the device runs the 12SL
algorithm in normal analysis mode and you see the
lower-acuity statements.
Default value: Disabled
Reason
Statements
Enable or disable
this setting.
You can select this setting only if you enable the
Screening Mode.
If this setting is enabled, reason statements generate or
display on the report when you perform the following:
• View a report in the preview and review screens
• View a stored report in the review screen
• Print a report
• Send a report to a configured destination.
If this setting is disabled, no reason statements
generate or display on the report.
Default value: Disabled
3. Click Save.
Configure Display Formats of ECG and Rhythm Leads
Table 35: Default Lead Formats for Each Lead Set
Name No. of
Leads
Lead
Format
Leads Default Auto-
rhythm
Digital
Rhythm
Adult 12 Lead
All
Leads
12 4x3 CH1 to CH12: I, II, III, aVR, aVL, aVF, V1,
V2, V3, V4, V5, V6
Yes No Yes
5864335-001-1  MAC