Configure Settings
Field Action Description
Password Enter the
password for the
user according
to the password
rules listed in
the Description
column.
Each character in the password displays an asterisk (*).
If the password rules are not met, the Password field
display a red box and relevant error messages.
Allowed values:
• User-defined value up to 126 characters
• Minimum number of characters and type of
characters allowed is set in the Security settings
screen. See Configure Security on page 201.
No default value.
NOTE:
If a local user forgets the user password, a user
with the User Management privilege can change
the password for the user account in the Users
settings screen. The local user can log into the
device with the changed password.
Confirm
Password
Enter the exact
duplicate entry
of the password
entered in the
Password field
Each character in the password displays an asterisk (*).
If there is a mismatch with the password entered in this
field and the Password field, the Confirm Password
field displays a red box. Re-enter the password to match
the Password field.
No default value.
Technician ID Enter the
Technician ID
associated with
the user.
This field can be blank.
No default value.
Allowed values:
• a to z
• A to Z
• 0 to 9
• All special characters
User-defined value up to 20 characters.
Password
Never Expires
Enable or disable
this setting.
• If this setting is enabled, the password of this user
does not expire, even if a duration for password
expiration for all users of this device is set in the
Password Expiration Duration field in the Security
settings screen.
• If this setting is disabled, the password of this user
expires when the time of the password exceeds
the duration for password expiration set in the
Password Expiration Duration field in the Security
settings screen.
Default value: Disabled
5864335-001-1  MAC
â„¢
5 A4/MAC
â„¢
5 A5/MAC
â„¢
5 Lite 215

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 5 A4 and is the answer not in the manual?

GE MAC 5 A4 Specifications

General IconGeneral
BrandGE
ModelMAC 5 A4
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals