Configure Settings
2. Configure the fields as per the information in the table.
Table 91: Configure Standby Modes
Field Action Description
Wireless
Standby
Enable or disable
this setting.
If this setting is enabled:
• The wireless connection is put in standby when the
device is in standby mode.
• The wireless connection is restored to its previous
state when the device wakes up.
Default value: Disabled
Auto Standby Enable or disable
this setting.
If this setting is enabled, the device is automatically put
on standby after a configured duration of inactivity.
Default value: Enabled
Auto Standby
Timeout (min)
Enter the duration
of inactivity, in
minutes.
This field is enabled when the Auto Standby setting is
enabled.
After this duration of inactivity, the device is
automatically put on standby.
Default value: 15
Allowed values: 5 to 120
Auto
Shutdown
Enable or disable
this setting.
If this setting is enabled, the device is automatically
shutdown after a configured duration of inactivity.
Default value: Disabled
5864335-001-1  MAC
â„¢
5 A4/MAC
â„¢
5 A5/MAC
â„¢
5 Lite 263

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 5 A4 and is the answer not in the manual?

GE MAC 5 A4 Specifications

General IconGeneral
BrandGE
ModelMAC 5 A4
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals