Patient Information
If Then
Only one ADT record is found on the MUSE system System searches for the Patient ID order found in
that ADT record.
• If multiple orders are found on the system -
The orders show on the Orders tab in a filtered
list. Select the order you will to attach to the
patient test.
• If only one order is found on the system - The
order fills in the Patient Information screen, if
there is no mismatch with the patient data on
the screen.
• If no orders are found on the system - A
message displays in the notification area that
no matching remote orders are found.
• If remote order query is failed on the system
- A message displays in the notification area
that remote order query has failed.
No ADT records are found on the MUSE system A message displays in the notification area that no
ADT data found.
Remote ADT data query failed on the MUSE
system
A message displays in the notification area that
ADT query has failed.
When local or remote orders are found, and you try to attach the order:
If Then
The selected order matches the patient
demographics in the Patient Information screen
A warning message displays on the screen:
Are you sure you want to attach this order to the
current test?
Select Yes to attach the order, No to cancel the
action.
The selected order does not match the patient
demographics in the Patient Information screen
A warning message displays on the screen:
*Name or Patient ID Mismatch*
Are you sure you want to attach this order to the
current test?
Select Yes to attach the order, No to cancel the
action.
Orders and then ADT Query Workflow
The device first searches for local orders with the Patient ID.
If Then
Multiple local orders are found on the device The orders show on the Orders tab in a filtered
list. Select the order you will to attach to the
patient test.