OPERATION
153
Choose whether to run the sequence in the
program automatically, skip the sequence
or manually start the sequence.
On-Time: 0.1 ~ 60.0 seconds
Determines how long the sequence will run
for (seconds).
Off-Time: Off – 0.1 ~ 60.0 seconds
Sets how long the sequence will stay off for
(in seconds) between each sequence.
Assuming Short time is not set to OFF, Off-
Time will always run after On-Time.
Short Time: Off – 0.1 ~ On-Time (seconds)
Determines how long a short circuit will
last (seconds). However the shorting time
cannot be longer than the On-time. Short
Time will start at the same time as On-time.
P/F Time: Off – 0.1 ~ (On-Time+Off-Time)-0.1
(seconds)
The Pass(P)/Fail(F) Delay Time can be set
to 0.1 seconds less than the total test time.
The total test time is defined as:
On-Time + Off-Time (seconds)
If Go/NoGo is turned on but the pass fail
time is off, then Go/NoGo test will
continue, but there will not be a specified
pass/fail time window.
Short Channel: Off – 1~ 8 (CH1~CH8)
Each channel can be individually set to
simulate a short circuit (CH1~8) or can
have shorting disabled (Off). When Short
Channel is set to Off, the channel will
ignore the execution of Short-Time.