154
7. Repeat the above steps for all ten
sequences for the same (current)
program.
8. Press F3 (Save) to save all the
sequence data for all of the
program.
F3
Note
Note the program data is not yet saved into Setup
Memory. If you wish to save the Program to Setup
Memory see page 248.
9. To recall the Default Program
settings, press F4.
F4
Note
If the Default is recalled, all data will be lost. This
does not include the internal Setup Memory. To see
the default settings, see page 302.
Below is a timing diagram of a single step in a
program.
On-Time Off-Time
Step test time
Start of Step End of Step
P/F End Test Time (fixed)
P/F Time
0.04s
0.06s
Short-Time
P/F Start Test Time (fixed)