PEL-2000A Series User Manual
174
Last Current(Last_C)
Sets the final current value
after OCP has been
tripped. This is the steady-
state current draw after the
OCP has been tripped.
Step Time(Step_T)
Sets the execution time of
each step. (50mS to 1600S)
Delay Time(Delay)
The OCP testing delay
time. Sets the how long to
delay starting the test after
the Load On key has been
pressed. (0 ~ 160S)
Trig Voltage(Trig_V)
Sets the voltage trigger
level needed see whether
the power supply OCP has
been triggered.
When the power supply
OCP has been triggered, its
voltage output will drop.
The voltage trigger level is
used to test to see if the
voltage output has been
drop.
Keep Time(Keep_T)
Set the how long to enter
the Last Current after
detect the OCP.(0~160S)
Note
This mode can only be used under CC mode.
1. Press the FUNC key, F4 (OCP) to
enter the OCP Test Automation
menu.
FUNC
F4