STATIC MODE
CONSTANT CURRENT MODE
Operating Range
0~2A
0~20A
Setting Range
0~2.04A
0~20.4A
Resolution
0.1mA
1mA
Accuracy
±(0.1%set + 0.1%F.S.
*1
)
±(0.1%set + 0.2%F.S).
CONSTANT RESISTANCE MODE
Operating Range
0.075Ω~300Ω(100W/16V)
3.75Ω~15kΩ(100W/80V)
Setting Range
0.075Ω~300Ω(100W/16V)
3.75Ω~15kΩ(100W/80V)
Resolution
0.333ms(100W/16V)
6.667μs(100W/80V)
Accuracy
300Ω: ±(0.2%set + 0.1S)
15kΩ: ±(0.1%set + 0.01S)
CONSTANT VOLTAGE + CONSTANT CURRENT MODE
Operating Range
1~16V
1~80V
Setting Range
0~16.32V
0~81.6V
Resolution
0.4mV
2mV
Accuracy
±(0.05%set + 0.1%F.S.)
Current Setting
Range
0~2.04A
0~20.4A
Resolution
0.1mA
1mA
Accuracy
±(0.1%set + 0.1%F.S.
*1
)
±(0.1%set + 0.2%F.S.
*1
)
CONSTANT POWER + CONSTANT CURRENT MODE
Operating Range
1~10W
1~100W
Setting Range
0~10.2W
0~102W
Resolution
1mW
10mW
Accuracy
±(0.5%set + 0.5%F.S.
*1
)
±(0.5%set + 0.5%F.S.)
Current Setting
Range
0~2.04A
0~20.4A
Resolution
0.1mA
1mA
Accuracy
±(0.1%set + 0.1%F.S.
*1
)
±(0.1%set + 0.2%F.S.
*1
)
NOTE : *1: F.S. = Full scale of H Range
DYNAMIC MODE
T1&T2
0.025ms ~ 10ms / Res: 1μs
10ms ~ 30s / Res: 1ms
Accuracy
1us / 1ms ± 100ppm
CONSTANT CURRENT MODE
Slew Rate
0.32 ~ 80mA/μs
3.2 ~ 800mA/μs

Table of Contents