GETTING STARTED
33
Voltage
Power
Load time
LOAD
Activates the load for the active
channel. (Right or Left)(A or B)
L
V sense
80V DC
MAX
The voltage sense terminals are
used when precise measurement is
needed. V Sense terminals are used
to compensate for voltage drops
across the main terminals caused
by the resistance of the load wires.
It is automatically activated when
connected to a DUT.
The voltage difference between the
DUT and load voltage should not
exceed 2V, otherwise you will get
incorrect measurement for the
voltage. (applicable to all models)
V sense
R
80V DC
MAX
L
80V DC
MAX
The terminals for both the left and
right side of a load can draw
differing amounts depending on
the load module specifications.
For 2 channel load modules, the left
terminals are used for the 1
st
channel and the right terminals are
used for the 2
nd
channel.
On single channel load modules,
the left terminals are the lower (-)
potential terminals, whilst the right
terminals are the positive (+)
potential terminals.
MAX
80V DC
R