OPERATING DESCRIPTION
79
If more than one Sequence is programmed on the
mainframe, the On End Of Seq. function will hold
the load current (of the selected sequence) to a
designated value until all the other sequences
have finished running.
End Of Seq. Example
Time
End Of Seq. = 1 Amps
End of
Sequences
End Of Seq. = OFF
Seq. 1
Seq. 2
Seq. 3
In the example above, Seq. 1 will hold the load
current at 1A at the end of its sequence until the
last sequence has finished. Seq. 2 is the longest
sequence, and as such the End Of Seq. setting is
not applicable. Seq. 3 is turned off after its
sequence has finished (0 amps).
The Trigger Out function allows a trigger
sequence signal to be output from a channel via
PIN 4 on Frame Link connector 1 when using
Sequences. The Trig Out function is used from the
Channel Duration menu.
Time
Trig
Out
Time
Value
Point1
Point2
Point3
Point4
Point5
Point6
CH1
Trigger Sequence