4-4 624074/07
4 Tests, calibrations and utilities
Required materials: Use the set up below appropriate to your
patient type. To ensure that the ventilator also functions
according to specifications on your patient, we recommend
that your test circuit be equivalent to the circuit used for
ventilation.
Adult patients • Breathing circuit, 19 mm ID with
22F connectors
• Flow sensor, pediatric/adult
• Demonstration lung, 2 l, with adult
ET tube between flow sensor and
lung
•CO
2
and SpO
2
sensors, adapters,
accessories
Pediatric patients • Breathing circuit, 15 mm ID with
22F connectors
• Flow sensor, pediatric/adult
• Demonstration lung, 0.5 l, with
pediatric ET tube between flow
sensor and lung
•CO
2
and SpO
2
sensors, adapters,
accessories
Neonatal patients • Breathing circuit, 10 mm ID with
10F connectors
• Flow sensor, infant
• Lung model, infant/neonatal, with
infant/neonatal ET tube between
flow sensor and lung model
(An IngMar neonatal lung model
is recommended)
•CO
2
and SpO
2
sensors, adapters,
accessories