Procedures:
Do or observe... Verify... Notes
1. Connect ventilator to
AC power, com-
pressed air and oxygen
supplies.
2. Assemble the patient
breathing circuit.
Breathing circuit is
assembled correctly.
Complete this task.
1. Switch power ON. When ventilator is
switched ON, an audio
alarm sounds and the
alarm lamp is red.
After the self-test is
passed the alarm lamp
turns red again.
The audio alarm
sounds briefly after the
device is switched ON.
1. 1. Open
System
and
Tests & calib
win-
dow (Figure 4-3).
2. 2. Select and run
Flow
Sensor
calibration.
3. Select and run
Tight-
ness
test.
4. Follow all system
prompts.
Each test result is
positive.
For details on running
these tests and
calibrations, refer to
Figure 4-3.
1. When necessary, run
O
2
cell calibra-
tion
.
2. Close window.
Each test result is
positive.