5-22 624074/07
5 Ventilator settings
Oxygen Oxygen concentration to be delivered.
Applies to all breaths in all modes.
In INTELLiVENT
®
-ASV Oxygen can be
set manual or automatic (see the
INTELLiVENT
®
-ASV manual).
NOTE:
Oxygen alarms are suppressed
for 30 seconds after an Oxygen
setting change.
21 to 100%
P-ASV limit Maximum pressure set by ASV
®
and
INTELLiVENT
®
-ASV, always equal to
Pmax-10. Thus P-max and P-ASV limit
co-vary.
PEEP+5 to
110 cmH2O
Patient Patient Profile. Adult, Pediat-
ric, Neonatal,
Male, Female
Patient height
(Adults only!)
Patient height. It determines the ideal
body weight (IBW), which is used in
ASV
®
, INTELLiVENT
®
-ASV and Intelli-
gent Panel calculations.
30 to 250 cm
(12 to
100 inch.)
Pause
1
Inspiratory pause or plateau, as a per-
centage of total breath cycle time.
After the required gas is delivered (after
the operator-set Vt is reached), gas
remains in the lungs and exhalation is
blocked during the Pause time. The use
of a Pause increases the residence time
of gas in the patient’s lungs.
Applies to volume-controlled man-
datory breaths, when the device is
configured in this way.
0 to 70% (of
cycle time)
Table 5-2. Control settings, mode additions, and ranges
Parameter Definition Range