624074/07 10-15
Check rate Low priority.
The chosen setting cannot be
obtained because of setting con-
flicts.
Confirm the proposed new
setting. Adjust other set-
tings as required.
Check
SpO
2
moni-
toring
Medium priority.
SpO
2
monitoring not available
Inspect probe attachment,
observe if probe site prop-
erly perfused
Check TI Low priority.
The chosen setting cannot be
obtained because of setting con-
flicts.
Confirm the proposed new
setting. Adjust other set-
tings as required.
Check %TI Low priority.
The chosen setting cannot be
obtained because of setting con-
flicts.
Confirm the proposed new
setting. Adjust other set-
tings as required.
Check trig-
ger
Low priority.
The trigger is OFF and the opera-
tor has attempted to activate a
mode allowing spontaneous
breathing. The ventilator
switches to the selected mode
and uses a pressure trigger of -3
cmH2O. It continues to alarm.
Verify the P-trigger setting
or switch the Flowtrigger
on.
Check Vt Low priority.
The chosen setting cannot be
obtained because of setting con-
flicts.
Confirm the proposed new
setting. Adjust other set-
tings as required.
Check
INTELLi-
VENT
®
PEEP limit
setting
HLI ≤ 10% and
VT/kg
≤
6 ml or (Ppeak exp pres-
sure
≤
10 cmH2O) and
PEEP ≥ 10 cmH2O
Verify if patient is haemody-
namic stable. Alarm is reset
when HLI limitation is deac-
tivated
CO
2
sensor
calibrated
OK
Operator Message. The CO
2
sen-
sor calibration was successful.
No action required.
Table 10-2. Alarms and other messages
Alarm Definition Complete these actions