624074/07 10-27
O
2
moni-
toring can-
not be
disabled
Operator message.
You attempt to disable O
2
moni-
toring during Heliox administra-
tion, this is not allowed.
No actions required
Oxygen-
ation
adjustment
OFF (no
SpO
2
)
Medium priority.
PEEP or oxygen automatic, SpO
2
invalid and needed since at least
60 seconds
Set PEEP and oxygen to
manual control
Oxygen-
ation
adjustment
OFF (no
SpO
2
)
Low priority.
PEEP or oxygen automatic, SpO
2
invalid and needed since at least
30 seconds
Set PEEP and oxygen to
manual control
Oxygen-
ation Con-
troller on
Limit
Low priority.
SpO
2
too low and PEEP and/or
oxygen can not be further
increased.
Observe patient.
Set oxygen and/or PEEP to
manual control or resolve
problem.
Oxygen set
to 100%
due to low
saturation
Medium priority.
The INTELLiVENT
®
-ASV has
increased oxygen to 100%.
Observe the patient.
Silence the alarm. Reset the
alarm by touching the
i-icon and/or alarm buffer.
Option key
already in
use
Operator Message. This option
has already been activated once
using this option activating code
(key).
Contact your HAMILTON
MEDICAL representative
Option key
invalid
Operator message. The option
code (key) you entered is not rec-
ognized.
Retry entry of valid option
code.
Option
unknown
Operator message. The option
you tried to activate is not avai-
lable for your ventilator’s soft-
ware version.
Retry entry of valid option
code.
Table 10-2. Alarms and other messages
Alarm Definition Complete these actions