EasyManua.ls Logo

Hamilton G5 - Page 310

Hamilton G5
594 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10-28 624074/07
10 Responding to alarms
Oxygen +
air supplies
failed
High priority.
Input pressures of both
oxygen and air < 1.9 bar
(190 kPa/27 psi), or the input
flow dropped below 40 l/min.
The device continues to ventilate
as long as there is sufficient reser-
voir pressure to maintain gas
flow. If the reservoir pressure
drops below 150 cmH
2
O, the
ventilator enters the ambient
state. This alarm cannot be
silenced.
Provide alternative ventila-
tion.
Check air and oxygen sup-
plies, or provide alternative
compressed air or oxygen
sources to the ventilator
(VENTILAIR
II
medical air
compressor or oxygen cylin-
der).
Oxygen +
heliox sup-
plies failed
High priority.
Input pressures of both oxygen
and heliox < 1.9 bar (190 kPa/
27 psi), or the input flow
dropped below 40 l/min. The
device continues to ventilate as
long as there is sufficient reser-
voir pressure to maintain gas
flow. If the reservoir pressure
drops below 150 cmH2O, the
ventilator enters the ambient
state. This alarm cannot be
silenced.
Provide alternative ventila-
tion. Check heliox and oxy-
gen supplies, or provide
alternative compressed air
or oxygen sources to the
ventilator (VENTILAIRII med-
ical air compressor or oxy-
gen cylinder).
Oxygen
supply
failed
High priority.
Input pressure of oxygen
< 1.9 bar (190 kPa/28 psi), or the
input flow dropped below 40 l/
min. The ventilator continues to
function at an oxygen concentra-
tion of 21% if the internal pres-
sure can be maintained. Under
these conditions, the low Oxygen
concentration alarm is disabled.
Observe the patient.
Inspect the oxygen supply.
Provide an alternative
source of oxygen, if neces-
sary.
Table 10-2. Alarms and other messages
Alarm Definition Complete these actions

Table of Contents

Related product manuals