624074/07 A-9
Pause
1
Adult/pediatric 0 to 70% (of
cycle time)
0% 5%
Infant/neonatal -- -- --
Pcontrol (pres-
sure control,
added to PEEP/
CPAP)
Adult/pediatric 5 to 100 cmH
2
O 15 cmH
2
O1 cmH
2
O
Infant/neonatal 3 to 50 cmH
2
O
2
15 cmH
2
O1 cmH
2
O
Peak Flow
1
Adult 1 to 180 l/min 54 l/min 1 l/min
Pediatric/infant/
neonatal
-- -- --
PEEP/CPAP
Adult/pediatric 0 to 50 cmH
2
O5 cmH
2
O1 cmH
2
O
Infant/neonatal 0 to 25 cmH
2
O5 cmH
2
O0.5 cmH
2
O for
0.1 cmH2O < 8 and
0.5 cmH2O > 8
P high 0 to 50 cmH
2
O 20 cmH
2
O1 cmH
2
O
P low
Adult/pediatric 0 to 50 cmH
2
O5 cmH
2
O1 cmH
2
O
Infant/neonatal 0 to 25 cmH
2
O5 cmH
2
O0.5 cmH
2
O for
0.1 cmH2O < 8 and
0.5 cmH2O > 8
Table A-5. Control setting ranges and resolutions
Setting Range Standard
setting
1
Resolution