A-10 624074/07
A Specifications
P-ramp (pressure ramp)
Adult:
P-CMV and
APVcmv
Other modes
50 to 200 ms
25 to 200 ms
50 ms
50 ms
25 ms
Pediatric/infant/
neonatal:
P-CMV and
APVcmv (as
applicable)
Other modes
50 to 200 ms
25 to 200 ms
100 ms
100 ms
25 ms
Psupport (pressure support, added to PEEP/CPAP or P low)
Adult/pediatric 0 to 100 cmH
2
O 15 cmH
2
O1 cmH
2
O
Infant/neonatal 0 to 50 cmH
2
O 15 cmH
2
O1 cmH
2
O
P-trigger (pressure trigger)
Adult/pediatric -0.5 to
10 cmH2O
(below PEEP/
CPAP)
-2 cmH2O 0.5 cmH2O
Infant/neonatal -0.1 to 5 cmH2O
(below PEEP/
CPAP)
2
-1 cmH2O 0.1 cmH2O for
< 2.0 cmH2O
0.5 cmH2O for
> 2.0 cmH2O
Table A-5. Control setting ranges and resolutions
Setting Range Standard
setting
1
Resolution