EasyManua.ls Logo

Hamilton G5 - Page 352

Hamilton G5
594 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A-10 624074/07
A Specifications
P-ramp (pressure ramp)
Adult:
P-CMV and
APVcmv
Other modes
50 to 200 ms
25 to 200 ms
50 ms
50 ms
25 ms
Pediatric/infant/
neonatal:
P-CMV and
APVcmv (as
applicable)
Other modes
50 to 200 ms
25 to 200 ms
100 ms
100 ms
25 ms
Psupport (pressure support, added to PEEP/CPAP or P low)
Adult/pediatric 0 to 100 cmH
2
O 15 cmH
2
O1 cmH
2
O
Infant/neonatal 0 to 50 cmH
2
O 15 cmH
2
O1 cmH
2
O
P-trigger (pressure trigger)
Adult/pediatric -0.5 to
10 cmH2O
(below PEEP/
CPAP)
-2 cmH2O 0.5 cmH2O
Infant/neonatal -0.1 to 5 cmH2O
(below PEEP/
CPAP)
2
-1 cmH2O 0.1 cmH2O for
< 2.0 cmH2O
0.5 cmH2O for
> 2.0 cmH2O
Table A-5. Control setting ranges and resolutions
Setting Range Standard
setting
1
Resolution

Table of Contents