624074/07 A-11
Rate
Adult: (S)CMV,
P-CMV, APVcmv
5 to 120 b/min 15 b/min 1 b/min
SIMV, P-SIMV,
APVsimv, Duo-
PAP
1 to 60 b/min 15 b/min 1 b/min
Pediatric:
(S)CMV, P-CMV,
APVcmv
5 to 120 b/min 25 b/min 1 b/min
SIMV, P-SIMV,
APVsimv, Duo-
PAP
1 to 60 b/min 25 b/min 1 b/min
Infant/neonatal:
PCMV, APVcmv
5 to 150 b/min 30 b/min 1 b/min
P-SIMV, APVsimv,
DuoPAP
1 to 80 b/min
2
30 b/min 1 b/min
Sigh Enabled,
disabled
Disabled --
T high
Adult 0.1 to 30.0 s 1.3 s 0.05 < 1.5 s
0.1 s > 1.5 s
Pediatric 0.1 to 30.0 s 0.8 s 0.05 < 1.5 s
0.1 s > 1.5 s
Infant/neonatal 0.1 to 30.0 s 0.6 s 0.05 < 1.5 s
0.1 s > 1.5 s
Table A-5. Control setting ranges and resolutions
Setting Range Standard
setting
1
Resolution

Table of Contents