624074/07 A-13
T low
Adult 0.1 to 30.0 s 2.7 s 0.05 < 1.5 s
0.1 s > 1.5 s
Pediatric 0.1 to 30.0 s 1.6 s 0.05 < 1.5 s
0.1 s > 1.5 s
Infant/neonatal 0.1 to 30.0 s 1.3 s 0.05 < 1.5 s
0.1 s > 1.5 s
Tolerance Time
(Quick Wean)
Adult: 10 to
600 s
Pediatric: 10 to
300 s
Adult: 180 s
Pediatric: 30 s
1
TRC (tube resistance compensation)
Adult:
Tube type
Tube size
Compensate
ET tube, Trach
tube, or TRC off
5.0 to 10.0 mm
10 to 100%
TRC off
7.0 mm
80%
Not applicable
0.5 mm
10%
Pediatric:
Tube type
Tube size
Compensate
ET tube, Trach
tube, or TRC off
3.0 to 7.0 mm
10 to 100%
TRC off
4.0 mm
80%
Not applicable
0.5 mm
10%
Trigger
Adult P-trigger,
Flowtrigger,
Trigger off
P-trigger --
Infant/neonatal P-trigger,
Flowtrigger,
Trigger off
Flowtrigger --
Table A-5. Control setting ranges and resolutions
Setting Range Standard
setting
1
Resolution